詳細(xì)介紹
Model 35
(dc to 26.0 GHz, 1.25 Megohm/0.05pF)
(Can be mounted on any micropositioner for use with any probing system)
The Picoprobe Model 35 has been engineered to meet the stringent demands of advanced high frequency circuit designers. This probe combines full DC capability, rise/fall times of 14 ps, and a nominal loading input impedance of 1.25 megohms shunted by 0.05 pf. Signal attenuation is 10:1 with a 50 ohm oscilloscope input. Like our other active probes, the Model 35 achieves its bandwidth using just one probe point, which greatly simplifies internal node testing without sacrificing performance. An assortment of user replaceable probe tips are available for a variety of probing needs. The Model 35 is powered by our new PS-3 power supply.
Specifications
Model 35
Input Capacitance
0.05pF
Input Resistance
1.25 megohms
Rise/Fall Time
14 ps (3V pulse or less)
Frequency Response
dc to 26GHz
Operating Range
-6 to +6V
Linearity
2% (+/-3V)
Signal Attenuation
10 to 1 (50ohm input)
FIGURE 2
Picoprobe Model 35
Part
NumberTungsten Probing Wire
Shaft DiameterPoint Radius
35-1-5
5 micron
<0.1 micron
35-1-10
10 micron
<0.1 micron
35-1-22
22 micron
<1.0 micron
35-1-35
35 micron
<2.0 micron
FIGURE 3
* Requires the use of a Model 35 Picoprobe錕?? Power Supply, part # PS-3. Please specify 110V or 220V when ordering.
![]()
FIGURE 4
This plot shows the response of the Model 35 Picoprobe to a 0.25 volt, 25 picosecond rise time input pulse. Both the input pulse (Channel 2) and the Model 35 response (Channel 1) are displayed simultaneously on a dual channel 20 GHz sampling oscilloscope. The Model 35 probe tip was contacted to a 50 ohm impedance strip line that carried the input signal. The Model 35 used only a single contact and did not require a short ground connection
制藥網(wǎng)






制藥網(wǎng)
安仁县|
青阳县|
喀喇沁旗|
无锡市|
怀来县|
射阳县|
石柱|
泰兴市|
墨脱县|
哈密市|
新余市|
新晃|
凌源市|
昌都县|
界首市|
清徐县|
简阳市|
达拉特旗|
合作市|
井陉县|
依安县|
徐州市|
海城市|
辽宁省|
江津市|
济宁市|
板桥市|
岳普湖县|
保定市|
兴安盟|
新野县|
新郑市|
潮州市|
延津县|
治县。|
连城县|
荥阳市|
康保县|
墨脱县|
东兰县|
静乐县|